"Telah dilakukan penelitian optimasi citra radiografi dengan phantom rando laki-laki menggunakan sistem FCR type Capsula XL-2 Drypic 4000. Dilakukan juga pengukuran ESD menggunakan thermoluminescent dosimeter (TLD), untuk pemeriksaan kepala AP, cervical AP, thorax PA teknik kVp standar dan teknik kVp tinggi, dan pemeriksaan pelvis AP. Optimasi pembentukan citra dievaluasi berdasarkan panduan dari European Commission dengan kriteria penerimaannya, kondisi eksposi kVp dan mAs, ESD, kontras tinggi dan kontras rendah. Selain evaluasi visual citra untuk optimasi diperhatikan pula karakter incident exposure FCR yang dinyatakan dengan Sensitivity Value (S Value) dengan proses digitasi citra yang dapat dilihat pada tampilan image consule dan softwere ImageJ.
Uji fungsi pesawat sinar-X dilakukan sebelum pelaksanaan penelitian menurut panduan RCWA, dan sistem FCR menurut panduan AAPM dan KCare, dengan hasil, keduanya memenuhi standar yang disyaratkan. Hasil penelitian optimasi menunjukkan bahwa untuk pemeriksaan kepala AP optimasi terjadi pada kondisi eksposi 65 kVp 20 mAs dan ESD 2.67 mGy. Pemeriksaan cervical AP optimasi terjadi pada eksposi 55 kVp 16 mAs dan ESD 2.55 mGy. Untuk pemeriksaan thorax PA teknik kVp standar optimasi terjadi pada 50 kVp 10 mAs dan ESD 2.24 mGy, sedangkan untuk teknik kVp tinggi optimasi terjadi pada eksposi 100 kVp 1 mAs dan ESD 1.75 mGy. Untuk pemeriksaan pelvis AP optimasi terjadi pada eksposi 75 kVp 10 mAs dan ESD 2.24 mGy.

A research about a radiography image optimization using a male rando phantom by FCR type Capsula XL-2 Drypic 4000 system has been done. Along with ESD measurement using a thermoluminescent dosimeter (TLD), for examination of AP skull, AP cervical, PA thorax use standard kVp technique and high kVp technique, and AP pelvis. The optimization of image formation was evaluated based on guidance from European Commission with their acceptance criterian, the condition of kVp and mAs, ESD, high contrast and low contrast. Beside the image visual evaluation for optimization, the FCR incident exposure was also observed which is stated in Sensitivity Value (S Value) by image digitations process that can be seen at image console and imageJ software.
Function test of X-ray device was done before the research based on the RCWA guidance, and FCR system based on the AAPM and KCare guidance, the results, both of them meet the standard. The result of optimization research show that for AP skull examination optimum condition was when the expose 65 kVp, 20 mAs and ESD of 2.67 mGy. For examination of AP cervical optimum condition was when the expose 55 kVp, 20 mAs and ESD 2.67 mGy. For the PA thorax, the optimization of standard kVp technique was when 50 kVp 10 mAs and ESD 2.24 mGy, for the high kVp technique optimization was when expose 100 kVp 1 mAs and ESD 1.75 mGy. And for the AP pelvis, optimization was when 75 kVp 10 mAs and ESD 2.24 mGy.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2012
T30282
UI - Tesis Open  Universitas Indonesia Library
cover
Nunung Nuraeni
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2008
T39884
UI - Tesis Open  Universitas Indonesia Library
cover
Ida Bagus Gde Manuaba
"Telah dilakukan pengukuran entrance surface dose (ESD) secara langsung menggunakan thermoluminescent dosimeter (TLD) pada jenis penyinaran dada thorax (chest). Data diperoleh dari 71 pasien di tiga rumah sakit (A, B dan C) yang menggunakan sistem pencitraan computed radiography (CR) Kodak. ESD ditentukan pula dengan metoda kalkulasi berdasarkan beberapa parameter pada pengukuran secara langsung. Metoda pengukuran lainnya juga dilakukan dengan menggunakan TLD pada objek radiasi berupa phantom dada buatan sendiri.
Penelitian ini diawali dengan beberapa pengukuran parameter seperti kVp, HVL dan tube output untuk mengetahui performa pesawat sinar-X. Pengukuran ESD pada phantom dan penentuan ESD metoda kalkulasi dilakukan pada kondisi penyinaran yang dibuat sama dengan kondisi penyinaran untuk pengukuran ESD pasien. Hasil penelitian menunjukkan nilai ESD thorax (PA) rata-rata di tiga rumah sakit berbeda secara signifikan. Hasil penelitian ESD thorax (PA) mendapatkan nilai ?faktor konversi metoda phantom? sebesar 0.874 sementara nilai ESD hasil perhitungan tidak berbeda signifikan dengan ESD pasien dengan penyimpangan maksimum sebesar ± 12 %.

Entrance surface dose (ESD) measurements have been carried out for chest examinations by means of thermoluminescent dosimeter (TLD). Data were collected from 71 patients at three hospitals (A, B, and C) which have been provided with Kodak computed radiography (CR). Based on exposure measurement parameters, ESD from these examinations was also calculated. An alternative method of measurement has been done with home-made chest phantom.
This work was initiated with measurements of X-ray tube parameters (kVp, HVL, tube output) to check equipment performance. Phantom-based measurement was set with the same exposure conditions as for patient measurements. The same exposure condition was also used in the calculation method. The results indicate that the ESD averages from chest examination (PA) at three hospitals were vary widely. Phantom-based conversion factor for ESD chest PA was found to be 0.874, whereas ESD from calculation is not differ significantly from the patient-ESD with maximum deviation ± 12 %.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2010
T29008
UI - Tesis Open  Universitas Indonesia Library
cover
cover
Ika Bayuadi
"Pemeriksaan Radiografi Thorak Posteroanterior (PA) merupakan pemeriksaan terbanyak didalam radiodiagnostik, diperlukan optimasi dosis dan citra radiografi Thorak PA. Penelitian teknik optimasi dengan menggunakan teknik kV tinggi pada pemeriksaan Radiografi Thorak PA dengan menggunakan reseptor system Kodak CRperlu dilakukan. Pengambilan citra radiografi Thorak PA dilakukan dengan objek fantom thorak dan sample pasien, untuk mengeliminir penilaian subjektif diambil citra radiografi fantom TOR 18 FG dan TOR CDR. Pada saat pengambilan citra radiografi thorak PA dilakukan pengukuran dosis masuk permukaan (Entrance surface dose / ESD ) dengan mengunakan TLD. Evaluasi citra radiografi thorak PA menggunakan 'quality criteria' European Commission EUR 16260 EN (1996). European Guidelines On Quality Criteria For Diagnostic Radiographic Images. Evaluasi dosis dengan membandingkan dengan dosis referensi dari IAEA BSS 115.Hasil evaluasi dosis pada fantom thorak, dari tiga variasi faktor eksposi 66 kV 8 mA, 85 kV 6.3 mAs dan 109 kV 2.2 mAs, dosis paling kecil dihasilkan dari faktor eksposi 109 kV 2.2 mAs.
Hasil evaluasi citra pada TOR 18 FG dan TOR CDR didapatkan sensitifitas kontras rendah, sensitifitas kontras tinggi dan resolusi pada kondisi 109 kV 2.2 mAs lebih besar daripada kondisi 66 kV 8 mAs. Hasil evaluasi gambaran thorak PA dengan mengunakan kondisi eksposi 109 kV 2.2 mAs dibandingkan dengan kondisi eksposi 66 kV 8 mAs kontras pada jaringan yang memiliki perbedaan kerapatan yang besar akan terjadi penurunan kontras. Sedangkan pada jaringan yang memiliki perbedaan kerapatan yang relatif kecil atau sama akan menaikkan kontras.

The Posterior-Anterior (PA) Examination of the thorax is the most frequent radio-diagnostic procedure. Optimization of dose and image of the PA thoracic radiography is required.This research was conducted to determine the optimal of techniques using high kV technique on thoracic radiography PA examination by using the Kodak CR system receptor and the patient sample, to eliminate the subjective assessment of radiographic image taken TOR 18 FG and TOR CDR phantom. At the time of image acquisition PA Thorax radiographs were performed, entrance surface dose measurements (ESD) were made using the TLD. The evaluation of the thoracic radiographic image of the PA using the quality criteria European Commission EUR 16260 EN (1996) European Guidelines On Quality Criteria For Diagnostic radiographic Images were made. An evaluation of dose by comparing with a reference dose of the IAEA BSS 155 was conducted.In the results of dose evaluation in thoracic phantom of the three variations of exposure factor: 66 kV 8 mAs, 85 kV 6.3 mAs and 109 kV 2.2 mAs, the smallest dose resulted from 109 kV 2.2 mAs exposure factor.
The result of the evaluation on TOR 18 FG and TOR CDR obtained low contrast sensitivity. The contrast sensitivity and higher resolution on the condition of 109 kV 2.2 mAs were larger than the condition of 66 kV 8 mAs. The results of the evaluation of thoracic image of PA by using the condition of 109 kV 2.2 mAs were comparable to the conditions of 66 kV 8 mAs contrast to the tissue that has large density differences that will decrease the contrast. While on a tissue that has relatively small density difference, or the same will increase the contrast.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2011
S200
UI - Skripsi Open  Universitas Indonesia Library
cover
Sudradjat
"Teknologi penggunaan Computed Radiography (CR) dalam radiologi diagnostik telah berkembang pesat. Saat ini ada beberapa merk CR yang banyak digunakan di Indonesia yang salah satu diantaranya adalah merk Fuji. Dalam pengambilan data citra CR dikenal dengan istilah pre-prosessing dan post-prosessing. Data citra pada pre-prosessing merupakan data citra yang belum mengalami proses komputerisasi pengolahan citra pada CR Fuji. Salah satu kekurangan dalam dalam data citra pre-prosessing adalah adanya noise dan kekaburan yang dapat menurunkan kualitas citra. Telah dilakukan penelitian dengan menggunakan alat uji Leeds Test Object Type 18 (High Contrast) dan Test Object 20 (low contrast).. Citra yang diperoleh dari pre-prosessing dan post prosessing kemudian dibandingkan. Sistem filtrasi yang terdapat pada Matlab digunakan untuk tujuan meningkatkan kualitas citra yang diperoleh dari pre-prosessing CR Fuji. Analisis dilakukan dengan menghitung kontras dan nilai SNR terhadap citra yang telah difiltrasi. Sistem filtrasi yang digunakan adalah imadjust, imsharpen, imhisteq dan adapthisteq. Setelah itu aplikasi penggunaan filtrasi dilakukan dengan citra foto thoraks, abdomen dan kepala.
Hasil penelitian menunjukan filtrasi dengan imadjust terhadap Leeds Test Object Type 18 (High Contrast) telah meningkatkan kontras citra rata-rata sebesar 59,86 %, sedangkan filtrasi menggunakan imsharpen telah meningkatkan kontras citra rata-rata sebesar 59,29 %. Filtrasi terhadap leeds test TO 20 (low contrast) telah meningkatkan kontras rata-rata sebesar 10,14 %. Peningkatan kontras dan resolusi ditandai secara visual semakin terlihatnya kelompok 1 lingkaran 7. Selain meningkatkan kontras, filtrasi juga memberikan noise terhadap citra yang difiltrasi. Peningkatan noise yang terjadi pada citra setelah difiltrasi adalah rata-rata sebesar 11,22 %. Peningkatan noise tersebut ditandai dengan nilai SNR yang semakin menurun.

Use of technology Computed Radiography (CR) in diagnostic radiology has been growing rapidly. Currently there are several brands CR which is widely used in Indonesia, one of them is a trademark of Fuji. In taking the CR image data is known as pre-processing and post-processing. The image of pre-processing is the image that has not undergone a process of computerized processing on Fuji CR. One of the flaws in the image pre-processing is the presence of noise and fuzziness that can degrade image quality. Studies have been conducted using test equipment leeds test object type 18 (high contrast) and test object 20 (low contrast). The image obtained from the preprocessing and post processing, then compared. Filtration system contained in Matlab is used for the purpose of improving the quality of the image obtained from the pre-processing. The analysis was performed by calculating the contrast value and SNR value of the image that has been filtered. The filtration system used is imadjust, imsharpen, imhisteq and adapthisteq. After the application of the use of filtration is done to the image of chest x-ray, abdomen and skull.
The results studies showed filtration with imadjust against leeds test object type 18 (high contrast) has improved image contrast on average by 59.86 %, while filtration using imsharpen has improved image contrast on average by 59.29 %. Filtration of the leeds test TO 20 (low contrast) has improved the contrast on average by 10.14 %. Improved contrast and resolution marked visually more visibility of group 1 loop 7. In addition to improving contrast, filtration also provides noise to the image. Increased noise that occurs in the image after filtered was an average of 11.22 %. Increased noise is characterized by decreasing SNR value."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2015
T43861
UI - Tesis Membership  Universitas Indonesia Library
cover
Sawiyah
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2010
T29000
UI - Tesis Open  Universitas Indonesia Library
cover
Suryo Adi Ari Santosa, supervisor
"Indonesia telah mengadopsi tingkat panduan dosis yang direkomendasikan oleh IAEA BSS 115 untuk beberapa jenis pemeriksaan sejak 2003. Pengukuran dosis masuk permukaan (entrance surface dose /ESD) untuk kepala dan cervical spine dengan menggunakan TLD dilakukan atas 4 pesawat sinar-X pada 3 rumah sakit di Jakarta yang diberi kode 1D, 1F, 1H dan 3H dengan reseptor sistem Kodak CR (computed radiography). Pengukuran ESD juga dilakukan berdasarkan parameter eksposinya menggunakan metode kalkulasi. Juga dilakukan pengukuran ESD pada fantom kepala ANSI dengan tujuan untuk mendapatkan faktor konversi fantom-pasien yang akan berguna dalam pengumpulan data ESD tanpa pasien.
Pelaksanaan penelitian dimulai dari penentuan ukuran standar orang dewasa Indonesia, uji fungsi pesawat sinar-X, preparasi TLD, pengambilan data ESD pasien dan diakhiri dengan pembuatan dan verifikasi fantom kepala ANSI. Sebanyak 20 orang pasien dewasa menjalani pemeriksaan kepala dengan 2 atau 3 proyeksi (AP/PA, lateral dan waters), dan 17 pasien dewasa menjalani pemeriksaan cervical spine dengan 2 proyeksi (AP dan lateral). Pengukuran ESD dengan TLD dengan cara menempatkan TLD pada kulit pasien pada pusat lapangan radiasi.
Dari pengukuran ESD menggunakan TLD pada pasien didapatkan nilai ESD pada kuartil ke-3 sebesar 1.4 mGy mGy, 1.4 mGy, dan 2.1 mGy untuk pemeriksaan kepala AP/PA, lateral, dan waters, serta 1.3 mGy dan 1.4 mGy untuk pemeriksaan cervical spine AP dan lateral. Sedangkan dari metode kalkulasi didapatkan nilai yang lebih tinggi, yaitu 2.7 mGy, 1.8 mGy dan 5.2 mGy untuk pemeriksaan kepala AP/PA, lateral, dan waters, serta 1.5 mGy for AP and 2.3 mGy untuk pemeriksaan cervical spine AP dan lateral. Rasio antara ESD pasien dengan fantom yang disebut faktor konversi fantom didapatkan sebesar 0.64 dan 0.77 untuk skull AP/PA dan lateral.
Dalam penelitian ini, nilai ESD lebih dipengaruhi oleh kebiasaan radiografer dalam penentuan faktor eksposi yang akan digunakan. Meski demikian, nilai ESD yang didapatkan cenderung lebih kecil daripada nilai referensi yang ada. Selain itu penggunaan fantom kepala ANSI akan sangat bermanfaat dalam pengumpulan data ESD mengingat jumlah pemeriksaan kepala cukup jarang.

Indonesia has adopted radiation dose guidance levels recommended by IAEA BSS 115 for several diagnostic examinations since 2003. Measurements of entrance surface dose (ESD) for head and cervical spine using TLD (thermo luminescence dosimeter) were carried out with 4 X-ray machines at 3 hospitals in Jakarta which were coded by 1D, 1F, 1H and 3H with Kodak CR (computed radiography) receptor system. Based on exposure, ESD for these examinations were also determined. ESD measurements were also conducted on ANSI skull phantom aiming to get conversion factors of phantom to patient that will be useful in collecting ESD data without patients.
This study began with determining Indonesian adult standard size, following by X-ray machine performance test, TLDs preparation, patient ESD measurement and ended with making ANSI skull phantom and ESD verification. There 20 adult patients for head examination with 2 or 3 radiography projections (AP/PA, lateral and waters), and 17 adult patients for cervical examination. The ESD was measured by TLD (thermo lunminiscence dosimeter), placed on the patient's skin, at the center of radiation field.
From the patient's ESD measurement using TLDs, the 3rd quartile ESD values were 1.4 mGy mGy, 1.4 mGy, and 2.1 mGy for head examination of AP/PA, lateral, and waters, while for AP and lateral cervical spine were 1.3 mGy and 1.4 mGy. Whereas ESD values from the calculation method based on exposure were higher, i.e. AP/PA, lateral and waters head examinations were 2.7 mGy, 1.8 mGy, and 5.2 mGy respectively, and for AP and lateral cervical spine examination were 1.5 mGy for AP and 2.3 mGy. The ratio between patient ESD and phantom ESD, namely conversion factors, were 0.641 and 0.765 for AP/PA and lateral skull examination.
In this study, ESD values found were more affected by technical habit of the radiographer in setting the exposure condition. In spite of that fact, it was found that these ESD values were still lower than references. Besides that, the usage of inhouse ANSI skull phantom will be beneficial for collecting ESD data considering the low frequency of head examinations.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2010
T29007
UI - Tesis Open  Universitas Indonesia Library
cover
Annisa Dwinda Fildzah Hani
"Penelitian ini bertujuan untuk mendapatkan kombinasi parameter eksposi optimum pada pemeriksaan sinar-x diagnostik menggunakan Computed Radiography (CR). Kombinasi faktor eksposi yang diuji berada dalam rentang 55 kVp-66 kVp dan 15 mAs-24 mAs untuk toraks, 81 kVp-102 kVp dan 8 mAs-20 mAs untuk abdomen, serta filter tambahan 0 mmAl, 1 mm Al + 0.1 mm Cu, 1 mm Al + 0.2 mm Cu, dan 2 mm Al. Figure of Merit (FOM) sebagai rasio antara kuadrat Signal Difference to Noise Ratio (SDNR) dan dosis dipilih sebagai parameter uji, dengan parameter kualitas citra tambahan berupa Modulation Transfer Function (MTF) dan Contrast Consistency (CV). Meskipun didapatkan kombinasi dengan FOM tertinggi, hasil penelitian menunjukkan bahwa FOM tidak dapat digunakan sebagai parameter optimisasi tunggal dan penggunaannya harus disertai parameter lain. Karenanya, diperlukan penelitian lanjutan sebelum metode ini dapat diterapkan secara klinis.

This study aims to obtain an optimum combination of exposure parameters on diagnostic x-ray examinations using Computed Radiography (CR). The combination of exposure parameters tested were in the range of 55 kVp-66 kVp and 15 mAs-24 mAs for thorax, 81 kVp-102 kVp and 8 mAs-20 mAs for the abdomen, and additional filters 0 mmAl, 1 mm Al + 0.1 mm Cu, 1 mm Al + 0.2 mm Cu, and 2 mm Al. Figure of Merit (FOM) as a ratio between the squared Difference to Noise Ratio (SDNR) signal and the Entrance Surface Dose (ESD) was chosen as optimization parameter alongside with additional image quality parameters such as Modulation Transfer Function (MTF) and Contrast Consistency (CV). Although the combination with the highest FOM was obtained, the results showed that FOM cannot be used as a single optimization parameter and its use must be accompanied by other parameters. Therefore, further research is needed before this method can be applied clinically."
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2018
S-Pdf
UI - Skripsi Membership  Universitas Indonesia Library
cover
Icha Renisha Mulia
"Pada pesawat angiografi modern, dosis maupun DAP selama pemeriksaan ditampilkan pada panel kendali. Namun, dosis tersebut tidak secara langsung menunjukkan dosis entrans kulit, utamanya pada pusat lapangan radiasi. Dengan menggunakan perangkat lunak berbasis AndroidTM telah dilakukan estimasi dosis entrans kulit pasien berdasarkan hasil pengukuran dengan fantom polymethyl methacrylate PMMA. Dosis entrans kulit pasien yang dikalkulasi dengan mengacu pada Kerma at Reference Point Ka,r dan Dose Area Product DAP dilakukan dengan memasukkan koreksi geometrik akibat kemiringan gantri, atenuasi meja pasien, koefisien konversi kerma udara menjadi dosis serap, dan faktor hamburan balik. Hasil kalkulasi ESD merupakan dosis titik pada pusat lapangan radiasi, bukanlah dosis maksimum yang diterima kulit pasien.
Hasil kalkulasi dibandingkan dengan hasil pengukuran ESD pada simulasi dengan fantom PMMA. Penelitian dilakukan dengan pesawat angiografi Siemens Artis Zee dengan kondisi eksposi 69-87 kV dan filter tambahan 0,1 mmCu. Pengukuran dengan thermo-luminescent dosemeters TLD dilakukan pada proyeksi penyinaran posterior-anterior PA , left anterior oblique LAO , right anterior oblique RAO , cranial CRA dan caudal CAU . Diskrepansi ESD hasil kalkulasi dan ESD pengukuran memiliki rata-rata 0,66-5,25 untuk kalkulasi mengacu DAP dan 0,52-5,17 . untuk kalkulasi mengacu Ka,r.

During examination of interventional radiology, radiation dose as well as Dose Area Product DAP are shown on control panel in modern angiography devices. However, dose shown indirectly indicates entrance skin dose received by patient, especially dose at center point of radiation field. Using AndroidTM based software, estimation of entrance skin dose ESD has been done based on measurement result using polymethyl methacrylate PMMA phantom in examination simulation. Patient entrance skin dose calculated using Kerma at reference point Ka,r and DAP as input value, then corrected by geometrical factor, patient table attenuation, air kerma to dose conversion coefficient, and backscatter factor.
Calculation result then compared to measurement result of ESD on PMMA phantom. The study was performed using Siemens Artis Zee angiography with exposure condition of 69 87 kV and additional filtration 0,1 mmCu. ESD measurements were carried out with thermo luminescent dosemeters TLD , in the projection of posterior anterior PA , left anterior oblique LAO , right anterior oblique RAO , cranial CRA dan caudal CAU . Discrepancy between ESD calculation and ESD measurement ranged from 0,66 to 5,25 for calculation using DAP as reference, in the other hand calculation using Ka,r as reference has discrepancy ranged from 0,52 to 5,17.
"
Depok: Fakultas Matematika dan Ilmu Pengetahuan Alam Universitas Indonesia, 2017