ABSTRACTThis thesis is designed to get the instrumentation with high sensitivity.
Generally, this thesis presents several noteworthy characteristics. First, Colpitts
oscillator is used as measurement instrumentation for temperature and humidity
with frequency output varying which correlate with those parameters. Second,
AT-Tiny2313-20PU is employed as frequency counter interfaced to a computer
through an RS-232 port. Third, LabVIEW is employed as the software
development environment to communicate from computer to external hardware
for automatically uploading and processing data. The increasing sensitivity gotten
is from (29,9exp-3,45E-2T) C/0C to (-9,73E4/T)Hz/0C and from (1,83E-9) F/%RH to
(-48,3E2)Hz/%RH for temperature and humidity respectively.